Home       Schedule a Call

National Supplier of Nurse Call & Life Safety Systems

Rauland Responder 4000 R4K Power Supply Pre-Owned

SKU
R4KPR400-REF
$185.00

This item is a for sale is a Rauland Responder 4000 R4K Power Supply Pre-Owned


OR
  • INFO
  • SPECS
  • RELATED
  • Rauland Responder 4000 R4K Power Supply Pre-Owned

    View Other Rauland Nurse Call REPLACEMENT Parts

    View Rauland Nurse Call REPAIR Options

    The R4KPR400 provides 15.5V DC power for localResponder 4000 components including consoles, annunciatepanels, patient stations, corridor lights, termination boards,K-bus, and X-bus devices. As a Responder 4000 system growsadditional power supplies can be added when needed (referto power calculation worksheets to determine the proper numberof R4KPR400s). 

    The optional R4KBK400 works in conjunction with anR4KPR400 to support emergency backup power for up to ten(10) minutes. The R4KPR400 has a modular connection to link itto the R4KBK400. One (1) R4KBK400 is required per R4KPR400when designing for emergency battery backup.

    Features

    • Self-contained unit
    • Status LED indicators
    • Three (3) individually energy limited outputs
    • Power connectors for termination board and K-bus components
    • Meets Class 2 energy limited • Meets Class 2 energy limited requirements
    • Power failure output with optional • Power failure output with optional R4KBK400 — Battery Backup Kit

    Specifications

    • Power Required: 120V AC to 250V AC @ 2.3A
    • Weight: 3.75 lbs. (1.70 kg)
    • Size: W: 11.43" (28.6 cm) H: 3.91" (9.9 cm) D: 3.25" (8.3 cm)
    • Housing &Finish: Black metal enclosure w/white nomenclature
    • Controls: One (1) AC On/Off switch One (1) AC status LED One (1) DC status LED
    • Environmental Parameters: 32°F (0°C) to 85°F (30°C) Relative Humidity – 0 to 85%
    • Power Output: Regulated 15.5V DC @ 3.5A total Output #1- 2.5A max output Output #2&3(shared) 1A max output
    • Heat Output: 25W or 85 Btu/hr max (@ 3.5A full load)
    • Terminations: Three (3) pair of power screw terminals (K-bus devices, termination board) One (1) AC power connector (IEC Universal receptacle) One (1) plug-on connector for battery backup unit
    • Backbox Requirements: Rauland NC2828 Terminal Cabinet
    • Certification: UL/C-UL 1069