  • Dukane Nurse Call 6000 ProCare Staff/Duty Station 4A2395A

    View Dukane Nurse Call REPAIR Options

    The System Monitor Function 

    When the system monitor is activated, all calls (normal, priority, and recall) will flash and tone at the monitoring stations. 
    Depending on how your system is configured, the system monitor can be in master mode, staff mode, or duty mode: 

    • In master mode, a monitoring station will flash and tone for any calls placed from stations covered by the same base master as the monitoring station. 
    • In staff mode, there are two stages: primary to primary, and primary to alternate. 
    • Note: This function does not operate if the ProCare 6000 system is equipped with version 5. XX video displays using MatchMaker ip . 
    • In the first stage, the call is indicated at a monitoring station that has the same primary staff assignment as the calling station. 
    • In the second stage, if the call is not responded to after a preprogrammed interval, the call is indicated at a monitoring station with a primary staff assignment that matches the calling stationís alternate staff assignment.
    • In duty mode, a monitoring station will flash and tone for any calls placed from stations in the same duty area as the monitoring station.
    • If your system is configured for the staff or duty monitor mode, the system activates staff or duty mode first. However, if all base masters covering the monitoring station are placed off duty, the system reverts to master mode

    Activating System Monitor at a Staff/Duty or Patient Station 

    Press the SYSTEM MONITOR button to activate the system monitor at a staff/duty or patient station. 

    When you do, this happens: 

    • The stationís SYSTEM MONITOR LED illuminates steadily. 
    • The red section of the corridor light illuminates (if so configured). 
    • When a call covered by this station is placed in the system, this station sounds a tone and flashes its SYSTEM MONITOR LED to indicate the call priority.
    • To silence a call tone, see Silencing Incoming Call Tones at a Monitoring Station 

    Notes: 

    Red presence is automatically registered when the SYSTEM MONITOR button is pressed. Registering the presence of a different level also activates the SYSTEM MONITOR function and illuminates the associated corridor light. 

    The system monitor function is always active on staff/duty stations configured as duty stations